Impedance Solutions

IEC 60601

IEC 60601 limits 100 µA maximum dc-leakage current through the body under normal conditions and 500 µA maximum under worst-case, single-fault conditions. An impedance circuit should not exceed that current.

Analog Devices AFE

Maxim AFE

  • Biopotential

    • ECG (waveform)
    • R-R (heart rate)
  • Bioimpedance

    • Resp (respiration)
    • GSR (galvanic skin response, stress level)
    • EDA (electro dermal activity)
  • Max 30001G ($10) for Biopotential and Bioimpedance, evaluation kit available ($113)

  • Maxim 30009 for Bioimpedance, evaluation kit available ($212)

AFE Summary

PART Features Resolution Remarks
AD5933 Single channel 3 wire 12bit, 1Msps, 100kHz excitation First Gen, 1024 point FFT, 1kOhm - 1MOhm
AD5940/41 2 channel, 4 wire 16bit, 800ksps, 200kHz Impedance spectroscopy
ADuC355 2 channel, 4 wire 16bit, 800ksps, 200kHz Impedance Spectroscopy, 26MHz ARM Cortex M3 MCU
MAX30002 single channel, BioZ, 2 or 4 wire 17bit
MAX30009 single channel, BioZ, 2 or 4 wire 16 kHz - 806 kHz
MAX30131/132/134 1-,2-,4- channel 12 bit
MAX30001G single channel, 2 or 4 wire BIOZ: 1-100kHz xx bit, ECG: xxbit Impedance and Potential (ECG)

References

For general reading: Analog Devices: Bio-Impedance Circuit Design for Body Worn Systems"

  1. Analog Devices AD5933 Evaluation Board, User Guide
  2. Munoz et a, 2022, A low-cost, portable, two-dimensional bioimpedance distribution estimation system based on the AD5933 impedance converterl
  3. AD5933 Datasheet
  4. Instructables: Body Composition
  5. Instructables: BIA 1
  6. Instructables: BIA 2
  7. Instructables: Thor 1
  8. Instructables: THor 2