The fundamental frequency of the high-priority signal must be between 150 and 1000Hz (it is now equal to 4075Hz).
Pulse duration is not respected (High priority alarm must consist of 5 pulses repeated twice (the current state has 2))
The text was updated successfully, but these errors were encountered: